2020 journal article

An Improved CMUT Structure Enabling Release and Collapse of the Plate in the Same Tx/Rx Cycle for Dual-Frequency Acoustic Angiography

IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 67(11), 2291–2302.

co-author countries: United States of America 🇺🇸
author keywords: Acoustics; Imaging; Transducers; Ultrasonic imaging; Finite element analysis; Electrodes; Frequency control; Contrast-enhanced ultrasound imaging; dual-mode operation; high-frequency (HF) receive; insulation spacers; low-frequency (LF) transmit; microbubbles
MeSH headings : Angiography / instrumentation; Angiography / methods; Contrast Media; Equipment Design; Finite Element Analysis; Microbubbles; Phantoms, Imaging; Transducers; Ultrasonography / instrumentation; Ultrasonography / methods
Source: ORCID
Added: June 21, 2021

This study demonstrates, in detail, the potential of using capacitive micromachined ultrasonic transducers (CMUTs) for acoustic angiography of the microvasculature. It is known that when ultrasound contrast agents (microbubbles) are excited with moderate acoustic pressure around their resonance (2-4 MHz), they produce higher order harmonics (greater than third harmonic) due to their nonlinear behavior. To date, the fundamental challenge has been the availability of a transducer that can generate the transmit signals to excite the microbubbles at low frequencies and, in the same cycle, confocally detect harmonics in the higher frequencies. We present a novel device structure and dual-mode operation of a CMUT that operates with a center frequency of 4.3 MHz and 150% bandwidth in the conventional mode for transmitting and a center frequency of 9.8 MHz and a 125.5% bandwidth in collapse mode for receiving. Output pressure of 1.7 MPapp is achieved on the surface of a single unfocused transducer. The mechanical index at the transducer surface is 0.56. FEM simulations are performed first to show the functionality of the proposed device, and then, the device fabrication is described in detail. Finally, we experimentally demonstrate the ability to detect the microbubble signals with good contrast, and the background reflection is adequately suppressed, indicating the feasibility of the presented approach for acoustic angiography.