2023 journal article

Energy calibration of germanium detectors for the MAJORANA DEMONSTRATOR

JOURNAL OF INSTRUMENTATION, 18(9).

By: I. Arnquist*, F. Avignone*, A. Barabash*, C. Barton*, K. Bhimani*, E. Blalock n, B. Bos*, M. Busch* ...

author keywords: Analysis and statistical methods; Double-beta decay detectors; Pattern recognition; cluster finding; calibration and fitting methods; Gamma detectors (scintillators, CZT, HPGe, HgI etc)
UN Sustainable Development Goal Categories
7. Affordable and Clean Energy (OpenAlex)
Source: Web Of Science
Added: January 2, 2024

Abstract The Majorana Demonstrator was a search for neutrinoless double-beta decay (0νββ) in the 76Ge isotope. It was staged at the 4850-foot level of the Sanford Underground Research Facility (SURF) in Lead, SD. The experiment consisted of 58 germanium detectors housed in a low background shield and was calibrated once per week by deploying a 228Th line source for 1 to 2 hours. The energy scale calibration determination for the detector array was automated using custom analysis tools. We describe the offline procedure for calibration of the Demonstrator germanium detectors, including the simultaneous fitting of multiple spectral peaks, estimation of energy scale uncertainties, and the automation of the calibration procedure.