2024 article

Multiangle Light Scattering as a Lentivirus Purification Process Analytical Technology

Sripada, S. A., Barbieri, E., Shastry, S., Wuestenhagen, E., Aldinger, A., Rammo, O., … Menegatti, S. (2024, May 28). ANALYTICAL CHEMISTRY.

By: S. Sripada n, E. Barbieri n, S. Shastry n, E. Wuestenhagen*, A. Aldinger*, O. Rammo*, M. Schulte*, M. Daniele n, S. Menegatti n

UN Sustainable Development Goal Categories
Source: Web Of Science
Added: June 11, 2024

The limited biomolecular and functional stability of lentiviral vectors (LVVs) for cell therapy poses the need for analytical tools that can monitor their titers and activity throughout the various steps of expression and purification. In this study, we describe a rapid (25 min) and reproducible (coefficient of variance ∼0.5–2%) method that leverages size exclusion chromatography coupled with multiangle light scattering detection (SEC-MALS) to determine size, purity, and particle count of LVVs purified from bioreactor harvests. The SEC-MALS data were corroborated by orthogonal methods, namely, dynamic light scattering (DLS) and transmission electron microscopy. The method was also evaluated for robustness in the range of 2.78 × 105–2.67 × 107 particles per sample. Notably, MALS-based particle counts correlated with the titer of infectious LVVs measured via transduction assays (R2 = 0.77). Using a combination of SEC-MALS and DLS, we discerned the effects of purification parameters on LVV quality, such as the separation between heterogeneous LV, which can facilitate critical decision-making in the biomanufacturing of gene and cell therapies.